FLUKE 9144-F-156-Dropship Automation

FLUKE 9144-F-156

SKU: DA-316123
Availability: 1 In Stock
Product Type: simple
Additional Information Part Number 9144-F-156 UPC 095969420501 P Specifications RTD measurement capability 4-wire RTD (2-,3-wire RTD w/ Jumpers only) Reference characterizations ITS-90, CVD, IEC-751, Resistance Built-in RTD thermometer readout accuracy NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C RTD connection 4 terminal input MA range Cal 4-22 mA, Spec...

Additional Information

Part Number
9144-F-156
UPC
095969420501

P Specifications

RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
RTD connection
4 terminal input
MA range
Cal 4-22 mA, Spec 4-24 mA
MA connection
2 terminal input
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
Reference probe connection
6-pin Din with Infocon Technology
TC connection
Small connectors
Voltage accuracy
0.025% of reading + 0.01 mV
Built-in electronics temperature Coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per°C
Reference resistance range
0 ohms to 400 ohms
Built-in reference thermometer readout accuracy (4-wire reference probe)1
±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C
Built-in mA readout accuracy
0.02% of reading + 2 mV
Reference measurement capability
4-wire
RTD resistance range
0 ohms to 400 ohms
RTD resistance accuracy2
25 ohms to 400 ohms: ±80 ppm of reading 0 ohms to 25 ohms: ±0.002 ohms
TC millivolt range
“10 mV to 75 mV
Loop power function
24 V DC loop power

Base Unit Specifications

Computer interface
RS-232 and 9930 Interface-it control software included
Heating time
15 min: 50°C to 660°C
Display
LCD,°C or°F user-selectable
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.015°C at 50°C ±0.025°C at 420°C ±0.035°C at 660°C
Stability
±0.03°C at 50°C ±0.04°C at 420°C±0.05°C at 660°C
Resolution
0.01°
Immersion (well) depth
150 mm (5.9 in)
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
Weight
7.7 kg (17 lb)
Display accuracy
±0.35°C at 50°C ±0.35°C at 420°C ±0.5°C at 660°C
Temperature range at 23°C
50°C to 660°C (122°F to 1220°F)
Hysteresis
0.1
Insert OD
24.4 mm (0.96 in)
Power requirements
100 V to 115 V (±10%), 50/60 Hz, 1400 W 230 V (±10%), 50/60 Hz, 1800 W
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Cooling time
25 min: 660°C to 100°C 35 min: 660°C to 50°C
Radial uniformity
±0.02°C at 50°C ±0.05°C at 420°C±0.10°C at 660°C
Axial uniformity at 40 mm (1.6 in)
±0.05°C at 50°C ±0.35°C at 420°C±0.5°C at 660°C

Returns Policy

You may return most new, unopened items within 30 days of delivery for a full refund. We'll also pay the return shipping costs if the return is a result of our error (you received an incorrect or defective item, etc.).

You should expect to receive your refund within four weeks of giving your package to the return shipper, however, in many cases you will receive a refund more quickly. This time period includes the transit time for us to receive your return from the shipper (5 to 10 business days), the time it takes us to process your return once we receive it (3 to 5 business days), and the time it takes your bank to process our refund request (5 to 10 business days).

If you need to return an item, simply login to your account, view the order using the "Complete Orders" link under the My Account menu and click the Return Item(s) button. We'll notify you via e-mail of your refund once we've received and processed the returned item.

Shipping

We can ship to virtually any address in the world. Note that there are restrictions on some products, and some products cannot be shipped to international destinations.

When you place an order, we will estimate shipping and delivery dates for you based on the availability of your items and the shipping options you choose. Depending on the shipping provider you choose, shipping date estimates may appear on the shipping quotes page.

Please also note that the shipping rates for many items we sell are weight-based. The weight of any such item can be found on its detail page. To reflect the policies of the shipping companies we use, all weights will be rounded up to the next full pound.

Recently Viewed Products