Additional Information Part Number 9142-A-156 UPC 095969419420 P Specifications Reference resistance range 0 ohms to 400 ohms RTD resistance range 0 ohms to 400 ohms Loop power function 24 V DC loop power RTD measurement capability 4-wire RTD (2-,3-wire RTD w/ Jumpers only) Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C) ±0.005% of range per°C MA range Cal...

Additional Information

Part Number
9142-A-156
UPC
095969419420

P Specifications

Reference resistance range
0 ohms to 400 ohms
RTD resistance range
0 ohms to 400 ohms
Loop power function
24 V DC loop power
RTD measurement capability
4-wire RTD (2-,3-wire RTD w/ Jumpers only)
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
±0.005% of range per°C
MA range
Cal 4-22 mA, Spec 4-24 mA
Built-in RTD thermometer readout accuracy
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C
TC millivolt range
-10 mV to 75 mV
MA connection
2 terminal input
RTD resistance accuracy2
0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
Reference probe connection
6-pin Din with Infocon Technology
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C)
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
Built-in mA readout accuracy
0.02% of reading + 2 mV
Internal cold junction compensation accuracy
±0.35°C (ambient of 13°C to 33°C)
Built-in reference thermometer readout accuracy (4-Wire Reference Probe)1
±0.010°C at -25°C ±0.015°C at 0°C±0.020°C at 50°C±0.025°C at 150°C±0.030°C at 200°C±0.040°C at 350°C±0.050°C at 420°C ±0.070°C at 660°C
Reference resistance accuracy2
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
TC connection
Small connectors
Built-in TC thermometer readout accuracy
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C
Reference characterizations
ITS-90, CVD, IEC-751, Resistance
Voltage accuracy
0.025% of reading + 0.01 mV
Reference measurement capability
4-wire
RTD characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
RTD connection
4 terminal input

Base Unit Specifications

Radial uniformity
±0.01°C full range
Environmental conditions (for all specifications except temperature range)
13°C to 33°C
Operating conditions
0°C to 50°C, 0% to 90% RH (non-condensing)
Cooling time
15 min: 23°C to -25°C 25 min: 150°C to -23°C
Display
LCD,°C or°F user-selectable
Hysteresis
0.025
Computer interface
RS-232 and 9930 Interface-it control software included
Power requirements
100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W
Temperature range at 23°C
-25°C to 150°C (-13°F to 302°F)
Weight
8.16 kg (18 lb)
Display accuracy
±0.2°C full range
Insert OD
30 mm (1.18 in)
Resolution
0.01°
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.006°C full range
Stability
±0.01°C full range
Immersion (well) Depth
150 mm (5.9 in)
Axial uniformity at 40 mm (1.6 in)
±0.05°C full range
Heating time
16 min: 23°C to 140°C 23 min: 23°C to 150°C 25 min: -25°C to 150°C
Size (H x W x D)
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)

Returns Policy

You may return most new, unopened items within 30 days of delivery for a full refund. We'll also pay the return shipping costs if the return is a result of our error (you received an incorrect or defective item, etc.).

You should expect to receive your refund within four weeks of giving your package to the return shipper, however, in many cases you will receive a refund more quickly. This time period includes the transit time for us to receive your return from the shipper (5 to 10 business days), the time it takes us to process your return once we receive it (3 to 5 business days), and the time it takes your bank to process our refund request (5 to 10 business days).

If you need to return an item, simply login to your account, view the order using the "Complete Orders" link under the My Account menu and click the Return Item(s) button. We'll notify you via e-mail of your refund once we've received and processed the returned item.

Shipping

We can ship to virtually any address in the world. Note that there are restrictions on some products, and some products cannot be shipped to international destinations.

When you place an order, we will estimate shipping and delivery dates for you based on the availability of your items and the shipping options you choose. Depending on the shipping provider you choose, shipping date estimates may appear on the shipping quotes page.

Please also note that the shipping rates for many items we sell are weight-based. The weight of any such item can be found on its detail page. To reflect the policies of the shipping companies we use, all weights will be rounded up to the next full pound.

Recently Viewed Products